Magnetocardiography Measurement Based on Optically Pumped Magnetometer and Multi-Branch Parallel Denoising Autoencoder
2024-07-31
会议录名称2024 ACADEMIC CONFERENCE OF CHINA INSTRUMENT AND CONTROL SOCIETY (ACCIS)
页码640-644
DOI10.1109/ACCIS62068.2024.10948626
摘要Magnetocardiography (MCG) is the measurement of the magnetic field generated by electrical activity in the heart using high-sensitivity magnetometers, which is a non-contact and non-invasive method for detecting early signs of myocardial ischemia. This paper presents a novel multi-view MCG system that utilizes two channels of optically pumped magnetometers (OPMs) with high-sensitivity to measure the magnetic field from different positions and directions. The OPMs are developed in our laboratory with small dimensions and high sensitivity, which operates in the spin exchange relaxation free range and has noise spectrum about 30 fT/Hz1/2. A MCG system including a magnetic shielding room, a magnetic field compensation system and a sensor system has been developed and tested. Then a multi-branch parallel denoising autoencoder is developed to denoise the signals. The results show that the system can capture the distinct cardiac waveforms. The multi-view measurements provide complementary information for the magnetic source localization and quantification.
会议录编者/会议主办者Chinese Instrument and Control Society (CIS)
关键词Electromagnetic induction Frequency shift keying Heart Image acquisition Image analysis Image thinning Magnetic field measurement Magnetic shielding Magnetometers Signal denoising Structural health monitoring Auto encoders De-noising Electrical activities High sensitivity Magnetic-field Measurement-based Measurements of Multi-branch parallel denoising encode Optically pumped Optically pumped magnetometer
会议名称2024 Academic Conference of China Instrument and Control Society, ACCIS 2024
会议地点Chengdu, China
会议日期28-31 July 2024
URL查看原文
收录类别EI
语种英语
出版者Institute of Electrical and Electronics Engineers Inc.
EI入藏号20251718309650
EI主题词Magnetocardiography
EI分类号101.3 Tissue Engineering ; 102.1 Medicine ; 103 Biology ; 214.2 Non-mechanical Properties of Materials ; 215.2 Testing of Non-mechanical Properties of Materials ; 701.1 Electricity: Basic Concepts and Phenomena ; 701.2 Magnetism: Basic Concepts and Phenomena ; 716.1 Information Theory and Signal Processing ; 746 Imaging Techniques ; 941.4 Magnetic Variables Measurements ; 942.1.5 Magnetic Instruments ; 1106 Computer Software, Data Handling and Applications ; 1106.3.1 Image Processing
原始文献类型Conference article (CA)
来源库IEEE
文献类型会议论文
条目标识符https://kms.shanghaitech.edu.cn/handle/2MSLDSTB/496955
专题信息科学与技术学院
信息科学与技术学院_PI研究组_叶朝锋组
信息科学与技术学院_硕士生
信息科学与技术学院_博士生
作者单位
1.School of Information Science and Technology, ShanghaiTech University, Shanghai, China
2.Shanghai Institute of Microsystem and Information Technology, Chinese Academy of Sciences, Shanghai, China
3.School of Information Science and Engineering, Northeastern University, Shengyang, China
第一作者单位信息科学与技术学院
第一作者的第一单位信息科学与技术学院
推荐引用方式
GB/T 7714
Jieya Chen,Chaofeng Ye,Limin Sun,et al. Magnetocardiography Measurement Based on Optically Pumped Magnetometer and Multi-Branch Parallel Denoising Autoencoder[C]//Chinese Instrument and Control Society (CIS):Institute of Electrical and Electronics Engineers Inc.,2024:640-644.
条目包含的文件
文件名称/大小 文献类型 版本类型 开放类型 使用许可
Magnetocardiography (730KB)会议论文 限制开放CC BY-NC-SA请求全文