摘要 | In this work, we present a prototype for the first time of using miniaturized aluminum nitride (AlN) micro-machined piezoelectric ultrasonic transducer (PMUT) arrays for non-invasive carotid artery monitoring. The designed hexagonal close-packed PMUT arrays device is in size of 3×3mm2, which contains four independent arrays. And any two of arrays can be used to accomplish pulse-echo method on skin tissue above carotid artery. The artery pulsation is successfully acquired with high signal-to-noise ratio and transformed to depth positioning image along time domain. Besides heart rate, the largest relative displacement of carotid artery wall to neck skin is also estimated via long time recording. |
会议录编者/会议主办者 | IEEE MEMS Technical Community |
关键词 | Artery monitoring PMUT arrays Portable medical ultrasonic system Pulsation wave Pulse-echo |
会议名称 | 36th IEEE International Conference on Micro Electro Mechanical Systems, MEMS 2023 |
会议地点 | deu,Munich |
会议日期 | 2023-01-15 |
学科门类 | Electronic, Optical and Magnetic Materials ; Condensed Matter Physics ; Mechanical Engineering ; Electrical and Electronic Engineering |
URL | 查看原文 |
收录类别 | EI ; SCOPUS |
语种 | 英语 |
出版者 | Institute of Electrical and Electronics Engineers Inc. |
EI入藏号 | 20231113731627 |
EI主题词 | Aluminum nitride |
EI分类号 | 704.2 Electric Equipment ; 712.1 Semiconducting Materials ; 716.1 Information Theory and Signal Processing ; 753.2 Ultrasonic Devices ; 804.1 Organic Compounds ; 804.2 Inorganic Compounds ; 921 Mathematics |
原始文献类型 | Conference Paper |
Scopus 记录号 | 2-s2.0-85149875611 |
来源库 | SCOPUS |
引用统计 | 正在获取...
|
文献类型 | 会议论文 |
条目标识符 | https://kms.shanghaitech.edu.cn/handle/2MSLDSTB/286544 |
专题 | 信息科学与技术学院_博士生 信息科学与技术学院_硕士生 |
作者单位 | 1.Shanghai Institute of Microsystem and Information Technology, Chinese Academy of Sciences 2.ShanghaiTech University, China |
推荐引用方式 GB/T 7714 | Sheng Wu,Kangfu Liu,Shuai Shao,et al. Non-Invasive Carotid Artery Monitoring by Using Aluminum Nitride Pmut Close-Packed Arrays[C]//IEEE MEMS Technical Community:Institute of Electrical and Electronics Engineers Inc.,2023:969-972. |
条目包含的文件 | ||||||
文件名称/大小 | 文献类型 | 版本类型 | 开放类型 | 使用许可 |
个性服务 |
查看访问统计 |
谷歌学术 |
谷歌学术中相似的文章 |
[Sheng Wu]的文章 |
[Kangfu Liu]的文章 |
[Shuai Shao]的文章 |
百度学术 |
百度学术中相似的文章 |
[Sheng Wu]的文章 |
[Kangfu Liu]的文章 |
[Shuai Shao]的文章 |
必应学术 |
必应学术中相似的文章 |
[Sheng Wu]的文章 |
[Kangfu Liu]的文章 |
[Shuai Shao]的文章 |
相关权益政策 |
暂无数据 |
收藏/分享 |
修改评论
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。